Molecules absorb infrared light when a vibration changes their dipole moment. This atlas connects the physics behind that sentence to the spectra you record: quantum oscillators, symmetry, the interferometer, sampling optics and band assignment. Every panel with a slider is a working model.
Simulated spectra built from Lorentzian and Gaussian bands at literature positions; intensities are schematic. Hover the trace or pick a band to see its assignment.
Spectroscopists quote wavenumbers because they are proportional to energy. The mid-IR covers fundamental vibrations; overtones and combinations spill into the near-IR; lattice modes, torsions and heavy-atom stretches live in the far-IR.
At room temperature \(k_BT/hc \approx 207\ \text{cm}^{-1}\), so vibrations above ~600 cm⁻¹ start almost entirely in \(v=0\). Hot bands matter for low-frequency modes and for heated samples.
A bond behaves, to first order, like a spring of force constant \(k\) joining two masses. Stiffer bonds and lighter atoms vibrate at higher wavenumber. This single equation explains most of the correlation chart.
Bond order sets \(k\): roughly 500 N m⁻¹ for single, 1000–1200 for double and 1600–1900 for triple bonds. That is why C–C ≈ 1000, C=C ≈ 1650 and C≡C ≈ 2200 cm⁻¹. Isotopic labelling (²H, ¹³C, ¹⁸O) is the cleanest experimental test of an assignment.
Real bonds dissociate. The Morse potential captures this: levels converge, overtones appear (weakly) at slightly less than integer multiples, and the dissociation energy follows from the spacing.
Electrical anharmonicity (non-linear \(\mu(Q)\)) and mechanical anharmonicity both give overtone intensity. Overtones and combinations are 10–100× weaker than fundamentals and dominate NIR spectroscopy.
χₑ = 0.0177 reproduces H–Cl (fundamental 2886 cm⁻¹ in the real molecule; ṽₑ here is rounded to 3000).
A molecule of \(N\) atoms has \(3N\) degrees of freedom. Removing translations and rotations leaves the normal modes: collective motions in which every atom oscillates at the same frequency and in phase.
The animation computes the dipole from point charges on the displaced atoms. Watch the arrow: if \(\boldsymbol{\mu}\) does not change during the motion, the mode is IR-silent. CO₂'s symmetric stretch is the classic case. It is Raman-active instead and sits in Fermi resonance with \(2\nu_2\), splitting into the 1285/1388 cm⁻¹ diad.
Whether a band appears, and how strong it is, follows from the transition dipole moment. Group theory tells you which modes survive before you do any calculation.
The dipole must change along the normal coordinate. Homonuclear diatomics (N₂, O₂) are IR-inactive.
Only adjacent levels couple through \(Q\). Anharmonicity relaxes this to \(\Delta v=\pm2,\pm3\) (overtones).
Polar bonds (C=O, O–H, C–O, Si–O) give strong bands; symmetric C=C and C≡C stretches are weak or absent. Unit check: 1 (D Å⁻¹)² u⁻¹ corresponds to 42.26 km mol⁻¹.
A mode is IR-active if its irreducible representation transforms like \(x\), \(y\) or \(z\); Raman-active if it transforms like a quadratic (\(x^2\), \(xy\), …).
In the gas phase each vibrational transition carries rotational structure. For a diatomic with \(\Delta J=\pm1\), the band splits into P and R branches around a gap at the band origin.
Polyatomics add a Q branch (\(\Delta J = 0\)) for perpendicular bands, as in CO₂ \(\nu_2\) at 667 cm⁻¹. Line widths give pressure broadening and collisional dynamics.
Absorbance is linear in concentration and path length. Transmittance is not. Quantitative work should stay roughly between A = 0.1 and 1; above that, stray light, detector non-linearity and finite resolution flatten the band.
Integrate bands rather than reading peak heights: integrated areas are less sensitive to resolution and to changes in band shape. Multivariate methods (CLS, PLS) extend this to overlapped mixtures.
A Michelson interferometer encodes all wavelengths at once. The detector records intensity against optical path difference \(\delta\); a cosine transform returns the spectrum. Resolution is set by how far the moving mirror travels.
\(D(\delta)\) is the apodization function. A boxcar (no apodization) gives a sinc instrument line shape with negative side lobes and FWHM \(0.603/\delta_{\max}\); triangular or Happ–Genzel windows suppress the lobes at the cost of a wider line.
Why FT wins. Fellgett (multiplex): all frequencies observed for the full scan time. Jacquinot (throughput): a circular aperture instead of a slit. Connes (precision): the He–Ne laser calibrates the wavenumber axis. Signal-to-noise grows as \(\sqrt{N_{\text{scans}}}\).
Attenuated total reflection is the default for solids and liquids today. The evanescent wave probes only a few micrometres, and that depth depends on wavelength, which distorts relative intensities compared with transmission.
| Technique | Best for |
|---|---|
| Transmission (KBr, cells) | Quantitation, gases, weak bands |
| ATR (diamond, Ge, ZnSe) | Neat solids, liquids, polymers |
| DRIFTS | Powders, catalysts, in-situ adsorption (Kubelka–Munk) |
| IRRAS / PM-IRRAS | Monolayers on metals (surface selection rule) |
| Microscopy / O-PTIR / AFM-IR | Spatial mapping, sub-µm to nm resolution |
Because \(d_p \propto \lambda\), ATR bands near 1000 cm⁻¹ look about three times stronger, relative to C–H stretches, than in transmission. Use your software's ATR correction before library searching. Near strong bands, anomalous dispersion in \(n_2\) also shifts peaks to lower wavenumber.
Localized vibrations of functional groups appear in predictable windows. Ranges below are typical for condensed phases; conjugation, ring strain, hydrogen bonding and substituent electronegativity move bands within and occasionally beyond them.
| Group / vibration | Range (cm⁻¹) | 4000 ← position → 400 | Intensity | Diagnostic note |
|---|
Intensity: s strong, m medium, w weak, v variable, br broad. “Show” marks the window on the spectrum at the top of the page.
Concepts that appear in research papers and thesis defences.
Homogeneous broadening (dephasing time \(T_2\)) is Lorentzian; inhomogeneous broadening from a distribution of environments is Gaussian. Their convolution is the Voigt profile.
A fundamental and an overtone of the same symmetry mix and share intensity. Examples: the CO₂ Raman diad, the aldehyde C–H doublet near 2820/2720 cm⁻¹.
The Badger–Bauer relation: stronger H-bonds red-shift, broaden and intensify X–H stretches. A band at 3200 cm⁻¹ instead of 3630 signals a strong interaction.
Kramers–Kronig links absorption \(\kappa\) to dispersion \(n\). It is used to convert specular reflectance into absorption-like spectra and to model ATR distortions.
Femtosecond pulse sequences spread the spectrum over pump and probe frequencies. Cross peaks reveal coupling; the time-dependent diagonal elongation tracks spectral diffusion through the frequency–frequency correlation function.
Harmonic DFT frequencies are typically scaled by about 0.95–0.97 depending on functional and basis set. Anharmonic (VPT2) calculations or MD dipole autocorrelation (above) capture overtones, band shapes and solvent effects.